Technical Specifications Of Analyzer Modules; Oxymat 7 Technical Specifications - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis, rack-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

Technical specifications

11.3 Technical specifications of analyzer modules

11.3
Technical specifications of analyzer modules
11.3.1

OXYMAT 7 technical specifications

Table 11-6
General
Information on the smallest span
Power consumption
Weight
Electrical safety
Table 11-7
Measuring ranges
Number of measuring ranges
Parameters can be as‐
signed in the measuring
ranges
Table 11-8
Gas inlet conditions
Sample gas pressure
Reference gas pressure
126
OXYMAT 7: General technical specifications
Warm-up phase
Operating phase ≤ 30 W
OXYMAT 7: Measuring ranges
Smallest possible spans
Largest possible span
OXYMAT 7: Gas inlet conditions
Standard devices with hoses
Tolerance within permitted pres‐
sure range
Standard devices with hoses and
ext. RG pump
Tolerance within permitted pres‐
sure range
Standard devices with pipes
Tolerance within permitted pres‐
sure range
High-pressure connection
Low-pressure version with external
reference gas pump
Without vibration compensation
See module nameplate or device certificate
≤ 60 W
< 5.5 kg (standard version)
In accordance with DIN EN 61010-1
Max. 4; parameters can be assigned
freely
0.5%, 1%, 2% or 5% O
100% O
500 to 1 500 hPa (absolute)
± 150 hPa
700 to 1 200 hPa (absolute)
± 150 hPa
500 to 3 000 hPa (absolute); short-term < 5000 hPa (abs.)
± 150 hPa
2 000 hPa above sample gas pressure
(within permitted reference gas pressure range 2500 to 5000
hPa, abs.)
100 hPa above sample gas pressure
Operating Instructions, 05/2015, A5E31874006-03
2
2
Rack-mounted device