Pepperl+Fuchs HID2891 Manual page 46

Frequency converter
Table of Contents

Advertisement

Frequency Converter HiD2891
6.5.4 Setting 0 V ... 5 V
V
5.125
5.0
0
0
0
Start value
At this setting, the start value of the measurement range is converted to 0 V and the end value to 5 V.
Intermediate values are converted proportionately.
Values below the start value cannot be evaluated (output 0 V). If the value exceeds the end value, the
output voltage increases linearly to a maximum of 5.125 V (approx. 102.5 % of the measurement
range). Further increases cannot be evaluated (output 5.125 V).
6.5.5 Setting 2 V ... 10 V unlimited
V
≈ 5.5
5.0
1.0
0
-25
At this setting, the start value of the measurement range is converted to 2 V and the end value to 10 V.
Intermediate values are converted proportionately.
If the value falls below the start value, the output voltage decreases linearly to a minimum of 0 V
(-25 % of the measurement range). Further decreases cannot be evaluated (output 0 V). If the value
exceeds the end value, the output voltage increases linearly to a maximum of 11 V
(approx. 112.5 % of the measurement range). Further increases cannot be evaluated (output approx.
11 V).
44
0
Start value
100
102.5
% Meas.
End value
range
≈ 112.5
100
% Meas.
End value
range

Advertisement