8 Maintenance
8.1 Care, cleaning, disinfection, and sterilization
54
8.1.4

Trackball holder

1. Rotate the cover ring counterclockwise and remove it.
2. Clean inner surface of cover ring (A) with ethanol (commercially
available cleaning alcohol).
3. Remove the ball.
4. Wipe out the calotte (spherical cap).
5. Insert the ball.
6. Fit the cover ring and turn it clockwise until it is firmly tightened.
NOTE
Setting the ease of action of the ball
For cover rings with various detent positions, the ease of action of the
ball can be set by selecting the corresponding detent position.
8.1.5

CEREC Omnicam

CAUTION
If the CEREC Omnicam accidentally falls down, check to make sure
that the camera window and coated sapphire glass are not damaged.
If the CEREC Omnicam has been damaged, it must no longer be
used on patients.
The CEREC Omnicam must be recalibrated.
NOTE
Do not under any circumstances sterilize the CEREC Omnicam or the
camera cable.
NOTE
The mirror sleeves cannot be sterilized in the autoclave.
NOTE
For markets where the RKI* guidelines are to be observed
The mirror sleeve is classified as a "semicritical medical device A"
according to RKI guidelines and therefore does not have to be
autoclavable.
*RKI=Robert Koch Institute, Berlin (Germany).