Replace Battery (Optional) - Dentsply Sirona CEREC AC Operating Instructions Manual

With cerec omnicam
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions CEREC AC
63 81 136 D3492
D3492.201.06.35.02    09.2019

Replace battery (optional)

8.6
A
Bottom door
B
Storage battery
1. Open the lower door on the back panel.
NOTE
Open with coin.
Use a coin to open the latch. Turn counterclockwise.
2. Remove the battery cover.
3. Unplug the battery connector.
4. Unscrew the fastening screw and remove the battery.
5. Insert the new battery into the battery compartment with the
fastening screw and screw it down.
6. Plug in the battery plug.
7. Attach the battery cover.
8. Put the door back in position and lock it.
8 Maintenance