Using the Probes

Functional Test

20
Use the following generic measurement procedure to ensure that
your probe is functioning properly.
1 Connect the probe to an oscilloscope channel input and ensure
the input impedance of the oscilloscope matches the output
impedance of the probe (50W).
2 Connect a grounding accessory to the probe and connect the
ground to the ground terminal on the oscilloscope.
3 Connect the probe tip to the oscilloscope's probe compensation
output.
4 Enable autoscale on the oscilloscope.
5 A square wave should now be displayed on the oscilloscope, if
the probe is functioning properly.
N2795/6A User's Guide