Chroma 19572 User Manual page 47

Ground bond tester
Table of Contents

Advertisement

It queries Output Meter value of the last step.
:SOURce:SAFEty:RESult:STEP<n>:JUDGment?
It queries judgment result code of the selected step.
:SOURce:SAFEty:RESult:STEP<n>:MMETerage?
It queries Measure Meter value of the selected step.
:SOURce:SAFEty:RESult:STEP<n>:OMETerage?
It queries Output Meter value of the selected step.
[:SOURce]:SAFEty:RESult:AREPort[:JUDGment][:MESsage]< boolean > | ON | OFF
This command is for setting if auto report the test result. When set as ON, the test is
completed then return "PASS" or "FAIL" string data. (RS232 interface only)
[:SOURce]:SAFEty:RESult:AREPort[:JUDGment][:MESsage]?
This command is for querying if auto report the test result, return 1 or 0 (RS232 interface only).
[:SOURce]:SAFEty:RESult:AREPort:OMETerage < boolean > | ON | OFF
This command is for setting if auto report the test result. When set as ON, the test is
completed then return the output current value of all STEPs. If some STEPs don't be
tested, there is without output value. These STEPs return is +9.910000E+37 (RS232
interface only).
[:SOURce]:SAFEty:RESult:AREPort:OMETerage ?
It queries the output current if auto report the test result, return 1 or 0 (RS232 interface only).
[:SOURce]:SAFEty:RESult:AREPort:MMETerage < boolean > | ON | OFF
It sets the measurement value if auto report the test result. When set as ON, the test is
completed then return the measurement value of all STEPs. If some STEPs don't be
tested, there is without measurement value. These STEPs return is +9.910000E+37
(RS232 interface only).
[:SOURce]:SAFEty:RESult:AREPort:MMETerage ?
It queries the measurement value if auto report the test result, return 1 or 0 (RS232
interface only).
[:SOURce]:SAFEty:ASTart:STATus?
This command is for querying GBSS MODE status, return character data ON|OFF.
[:SOURce]:SAFEty:STEP<n>:DELete
This command will delete STEP of <n> represented, the next STEP of <n> will fill it.
[:SOURce]:SAFEty:STEP<n>:SET?
This command is for querying all setting values in selected STEP.
[:SOURce]:SAFEty:STEP<n>:MODE?
This command is for querying MODE in selected STEP, it will return character data GB.
[:SOURce]:SAFEty:STEP<n>:GB[:LEVel]
This command is for setting selected STEP, ground resistance test needed current value,
the unit is ampere (A).
Remote Interface Description
< numeric value >
5-9

Hide quick links:

Advertisement

Table of Contents
loading