How Does One Recognize Devices With Ce Conformity Assessment; Calculator Settings: Heat And Cooling Calculator - INTEGRA Metering CALEC energy master Installation And Operating Instructions Manual

Hide thumbs Also See for CALEC energy master:
Table of Contents

Advertisement

17.2.3 How does one recognize devices with CE conformity assessment?

Devices with CE conformity assessment can be recognized by:
1. the CE mark on the type plate (see section 4.1),
2. the indicated values for the calibration date and calibration validity.
Go to page 1 of the main display (start window and
Without conformity assessment
ImpVal
1.000 L
Side Q
Cold side
Settings can be changed in programming
mode.

17.2.4 Calculator settings: Heat and cooling calculator

Warning!
conformity-assessed devices, the activation of programming mode invalidates the cali-
bration.
Parameters for the flow sensor
If a device is supplied as part of a measuring point (e.g. a combined heat meter), then the flow
sensor parameters have already been set correctly.
The corresponding parameters can be checked in the Basic settings / input 1 submenu (example
for pulse input Namur 200 Hz):
Input 1
Mod-No. 2
T-Nr..
82-10-11
Signal
ID Namur 200 Hz
Funct.
Pulse
Meas.var. Volume
Side Q
cold side
Status
0.000 Hz
ImpVal
1.000 L
...
FWV
V01.00.05
*: with the cooling meter application, Side Q = hot side is preferable (default setting).
®
CALEC
energy master installation and operating instructions
The following parameters can only be set in programming mode. With
Settable?
No
No
Yes
Signal type
Yes
Yes
Yes
No
Yes
No
).
With conformity assessment
ImpVal
1.000 L
Side Q
Cold side
E-Date.
30.06.2008
Valid
30.06.2013
Calibration-relevant settings must not be changed
otherwise the calibration will be invalidated.
Parameter
Select input with the
Module number
Terminal numbers
Function (status, pulse or frequency)
Measured variable (e.g. volume, mass...)
Installation side of flow sensor (hot/ cold side) *
Pulse frequency at input
Pulse value
Other parameters, not settable
Firmware version number of the module
keys
39

Advertisement

Table of Contents
loading

Table of Contents