Easystart Identification; Restriction Codes For Parameterization And Digital Setpoint; Save And Load Parameters With Usb Stick - Advanced Energy THYRO-AX Manual

Thyristor power controller
Hide thumbs Also See for THYRO-AX:
Table of Contents

Advertisement

Available languages

Available languages

60
Confirmation of the settings:
When Yes is pressed all settings are saved in the internal
EEPROM and the pulse inhibition is ended again. Pressing
the No button returns to the last screen.
The characters in brackets each form an initial for the
setting and are used for the EasyStart identification.

7.5 EASYSTART IDENTIFICATION

The EasyStart identification makes the comparison of the configurations
of multiple devices easier. Depending on the settings
selected a series of characters is generated which corres-
ponds to the selected settings. If additional settings have
been made outside of those covered by EasyStart, then
a + (plus symbol) is added at the end of the identification
code. This is an indication of additional alterations which
have not been created by EasyStart.
The EasyStart identification can be reached from the main
screen through pressing the actual value button multiple
times. The last line gives the identification code.
7.6 RESTRICTION CODES FOR PARAMETERIZATION
AND DIGITAL SETPOINT 2
The access via the touch display can be disabled and enabled for the
following contents:
- parameter alteration: setting menu is displayed or hidden (factory
setting: on)
- digital setpoint 2: DS2 is alterable via buttons following the screens of
the actual value view (factory setting: off).
If the actual value view is not already called up this can be reached via
Menu
ActVal. The screen for the DS2 follows the other actual value
screens and appears last. An alteration of the setpoint is possible with
the button + and -. The factory setting is for this value to be added to
the other setpoints and as such can be used as an offset.
Start
The functions can be hidden or displayed by entering the restriction
code under Menu
Code.
- DS2 restriction code: enable 234, disable 432 (restricted by default
settings).
- Parameter restriction code: enable 345, disable 543 (restricted by
default settings).

7.7 SAVE AND LOAD PARAMETERS WITH USB STICK

Parameters can be transferred from one device to another of the same
type by using a USB stick.
If a USB stick is inserted, a request will appear to set the pulse inhibiti-
on or not. Thereby the output power will be 0. This is a requirement for
loading and saving of parameters from or on the USB stick.
After loading, an option will appear to read parameters from the USB
stick (when parameters are available for the specific device type on the
USB stick and the device has not been locked by password protection
(s. chapter 7.6)) or to save the current parameters of Thyro-AX (on the
USB stick while existing parameters of the same device type will be
overwritten).
After reading, it is requested if the parameters should be saved perma-
nently on the device (EEPROM).
The USB overview can be seen when the USB stick is connected via its
according menu item.
On the market there are various USB sticks, which mainly can be
operate without any difficulties with Thyro-AX. Special sticks won't be
recognized which provide additional drives or need special drivers.