6.17 Connection to the Electronics Box CTR4000,
CTR5000, CTR5500, CTR6000, CTR6500,
CTR7000, CTR6500 HS, CTR7000 HS
The Leica DMI 3000 B is supplied without an
electronics box. The power supply is integrated
in the stand and a socket has been provided on
the back of the microscope to connect the trans-
mitted light illumination. The illuminated ON/OFF
switch is located on the stand.
Fig. 77
Rear view of electronics box, example: CTR6000
1
Mains socket
2
Earth screw
3
Type label
4
Port 12V 100W
for the lamp cable of the microscope
5
DL: reset-button
6
Port USB for communication with PC
7
Port Ext, IC interface
for additional external components
8
Port Trigger 2 for external synchronisation
9
Port Trigger 1 for external synchronisation
10 Port SmartMove
for Leica SmartMove control device
11 Port Microscope
for communication with the microscope
12 Port XY-Stage for motorized stage
13 Equipotential bonding symbol
14 Equipotential bonding
4
3
2
1
CTR 4000 Electronics Box
The Leica DMI 4000 B is supplied with the
CTR4000 electronics box. The power supply
for the microscope is located in this box. Two
sockets are located on the back of the CTR4000
electronics box for 12V/100W transmitted light
and 12V/100W incident light illuminators. The il-
luminated ON/OFF switch for the microscope is
located on the CTR4000 electronics box.
Fig. 78
Rear view of stand
1
RS232 ports
2
2 x USB
3
4 x EXT.
4
Connection for SmartMove
5
Electronic box connection
6
Condenser cable
7
Lamp power cable
5
6
7
8
9
10
11
12
1
13
2
14
6. Assembly
7
6
5
4
3
59