6. Assembly
6.10 Installation and Replacement of the
transmitted Light Lamps:

107 or 107/2 Lamp Housing

This lamp housing is used with a 12V 100W Halo-
gen Lamp, which is already mounted.
In case the lamp has to be removed:
Changing the 12 V 100 W halogen lamp
Caution!
Ensure that the lamp housing has been dis-
connected from the power supply. Unplug the
power plug and the power supply during as-
sembly.
Caution!
Light sources pose a potential irradiation risk
(glare, UV-radiation, IR-radiation). Therefore,
lamps have to be operated in closed hous-
ings and only after being mounted.
Fig. 50a
Lamp housing 107/2
Releasing the
fastening screw
46
• Remove the fastener screw on the housing
(Fig. 50a).
Caution!
The lamp and the lamp housing may still be
hot.
• Lift the housing off (Fig. 50b).
• Remove the lamp.
Caution!
Do not remove the new lamp's dust cover un-
til you have installed the lamp. Avoid fi nger-
prints on the lamp.
Fig. 50b
Removing housing
Fig. 50c
Lamp housing 107/2
opened
1
Mount with
halogen lamp
2
Collector
1
2