7.2 Switching on the Microscope

Leica DMI3000 B:
• Switch on the microscope's power at the On/
Off switch. The signal lamp is lit when the in-
strument is ready. (For the Leica DMI3000 B
please continue at 7.4. Function Buttons on the
Stand)
Leica DMI4000 B and Leica DMI6000 B:
• Switch on the power of the electronics box at
the On/Off switch (80.1). The signal lamp (80.2)
is lit green when the unit is ready. All motor ized
microscope components will then run through
an initialization phase.
Note:
If a PC is connected, switch on the electronics
box fi rst, and then the computer.
After the initialization (Fig. 81) is complete, the
LeicaScreen will display the microscope's cur-
rent settings (Fig. 82).
Fig. 80
front side Leica CTR6000
1
On/Off switch
2
Signal lamp
3
Warning: Observe safety notes!
If a component has not been installed correctly,
the LeicaScreen will display an error message.
See Troubleshooting chapter, → p. 106.
Components such as diaphragms, condensers,
light, and phase rings have been pre-centered
at the factory. It may be necessary to correct the
centering after the microscope has been trans-
ported and assembled.
Before performing the required steps, please fa-
miliarize yourself with the LeicaScreen and the
controls.
After turning on the gas discharge lamp, the
burner must be immediately adjusted. There-
fore, do not turn on the power supply unit
yet. First, work in transmitted light in order
to familiarize yourself with the microscope's
controls.
Fig. 81
LeicaScreen
Initialization
Fig. 82
LeicaScreen
after
Initialization
1
2
3
Caution!
7. Start-up
65