B2 All Available Commands - Dalsa Piranha 3 Series User Manual

8k to 12k line scan ccd cameras
Table of Contents

Advertisement

76
Message
Error 02:
Unrecognized
command >
Error 03: Incorrect
number of
parameters>
Error 04: Incorrect
parameter value>
Error 05: Command
unavailable in this
mod e>
Error 06: Timeout>
Error 07: Camera
settings not saved >
Error 08: Unable to
calibrate - tap
outsid e ROI>
Error 09: The
camera's
temperature exceed s
the specified
operating range>

B2 All Available Commands

As a qu ick reference, the follow ing table lists all of the com m and s available to the cam era
Parameters:
i = integer
u ser. For d etailed inform ation on u sing these com m and s, refer to Chapter 3.
t = tap id
Command
i = integer value
calibrate analog offset
f = real number
s = string
x1 = horizontal pixel start
number
x2 = horizontal pixel end
number
y1 = vertical pixel start
number
y2 = vertical pixel end
number
correction calibrate fpn
03-032-10216-06
All manuals and user guides at all-guides.com
Description
Command is not available in the curr ent access level or it is not a
valid command .
This response returned for
· Alpha received for numeric or vice versa
· Not an element of the set of possible values. E.g., Bau d Rate
· Outsid e the range limit
Command is valid at this level of access, but not effective. Eg line rate
w hen in smart Exsync mod e
Command not completed in time. Eg FPN/ PRNU calculation w hen
no external exsync is present.
Tried saving camera settings (rfs/ rus) but they cannot be saved .
Cannot calibrate a tap that is not part of the end of line statistics.
Ind icates that the camera has shut itself d ow n to prevent d amage
from further overheating.
Syntax
cao
ccf
Parameters
Description
t i
Calibrates the analog gain and averages
each tap's pixels w ithin the ROI to the
specified average target value.
t = tap selection, either 1 to number of
CCD taps, or 0 for all taps
i = target value in a range from 0 to
255DN (12-bit LSB)
Refer to Analog Signal Processing:
Setting Analog Gain and Offset for
d etails.
Start FPN coefficient calibration. Refer to
section 3.4.2 Analog and Digital Signal
Processing Chain for d etails.
Piranha 3 User Manual
Teledyne DALSA

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents