2> Sin/Cos Signal Check - Magnescale SR138-025R Instruction Manual

Scale set/unit
Table of Contents

Advertisement

Available languages

Available languages

<2> SIN/COS signal check
Note
• Be sure that the interface unit is turned off before attaching or
removing the probe of the oscilloscope.
• The check pin of the interface unit has little physical strength,
and so use an IC clip or other implement to connect.
(Recommended IC clip: SMG50 grabber clip by Tektronix,
etc.)
• Check that the interface unit is turned off.
• Refer to section for removing and attaching the cover of the
interface unit.
1 Connect the CH1 probe of the oscilloscope to the TP1 (SIN)
and TP4 (GND) check pins of the interface unit.
2 Connect the CH2 probe of the oscilloscope to the TP2 (COS)
and TP4 (GND) check pins of the interface unit.
26 (E)
Interface unit
CN2-5pin (Z)
Switch 1
(Reference point
POS1sw)
CL1 (ZG)
Switch 2
(Reference point
POS2sw)
TP4 (GND)
RV5 (REF DC)
RV6 (REF amplitude)
RV3 (COS DC)
RV4 (COS amplitude)
RV1 (SIN DC)
Signal adjustment
RV2 (SIN amplitude)
section-datailed view
Fig. 4-2
Oscilloscope
CH2
CH1
Probe
TP3 (REF)
TP2 (COS)
TP1 (SIN)