2> When The Scale Mounting Surface; Is A Casting Surface - Magnescale SR138-025R Instruction Manual

Scale set/unit
Table of Contents

Advertisement

Available languages

Available languages

<2> When the scale mounting surface is a
casting surface
Mount the scale part using mounting brackets
(1) When the scale mounting surface is a casting
surface and parallelism is not satisfied
Use mounting brackets and adjust so that the parallelism
specified in <1> may be satisfied.
The mounting brackets need only cover the length of at least
the feet of the scale part. However, the brackets must be long
enough for the parallelism of the right and left feet of the
scale part to be measurable.
If a problem occurs in the temperature tracking performance
with the machine, try to place the mounting brackets over the
entire length of the scale unit.
Parallelism measurement of side A
Pick tester
Parallelism measurement of side B
Fig. 3-6 Parallelism measurement of the mounting brackets
If the mounting bracket shape has an abutment surface,
adjust it so that the surface accuracy reading is within the
range shown in Section <1>. Use the procedure in (4) to
make adjustments if there is no abutment surface. Also, use
the procedure in (4) when the installation surface accuracy
cannot be obtained.
Mounting bracket
(E) 13

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents