SECTION 10: Error Messages, Warning Messages and General
Warning
Warning
Messages
General Warning:
Stabilization Failed:
Self-test Failed:
Revision 21_Sep_2015
The SQA-V equipment's built-in protection for the operator and the
environment is ONLY operational if the SQA-V is operated properly following
the manufacturer's specifications.
CAUTION: There is a risk of explosion or shorting if the SQA-V battery is
replaced by an incorrect type. Replacement batteries MUST be the same type
and manufacturer. Dispose of used batteries in accordance with the
manufacturer instructions.
Environmental condition for storage and transport: Recommended to store the
SQA-V at temperatures between 20°C -30°C.
Ensure there is no testing capillary in the measurement compartment.
Remove the SQA-V from sources of electronic noise and vibrations.
Clean measurement compartment (refer to Appendix).
Reboot the SQA-V without a testing capillary in the chamber:
Turn system OFF then back ON at the main switch on the rear panel.
Press the front panel ON/OFF key to begin Auto-Calibration/Stabilization.
Call technical support if failure recurs.
Ensure there is no testing capillary in the measurement compartment.
Remove the SQA-V from sources of electronic noise and vibrations.
Clean measurement compartment (refer to Appendix).
Reboot the SQA-V without a testing capillary in the chamber:
Turn the system OFF then back ON at the main switch on the rear panel.
Press the front panel ON/OFF key to begin Auto-Calibration/Stabilization.
29