User Guide Version 2.60 I-button WHO 5th
Method comparison:
SQA-V was compared to the microscope based on WHO 3
(Trial #1), 4
th
(Trial #2) and WHO 5
Sensitivity and Specificity were calculated using ROC
curves with the cutoffs based on the reference values of WHO
rd
th
th
3
, 4
and 5
guidelines (see Table #1).
Correlation coefficients of the SQA-V results to the manual
method are presented in the Table #2.
Precision: Inter-device (Tables #3) and intra-device (Table
#4) variations were compared to inter- and intra-operator
variability using Coefficients of Variation (CV, %). Duplicate
samples were assessed by two methods. The CVs
characterizing precision were calculated for multiple semen
parameters.
The POSTVASECTOMY test (Trial #3) compared three
assessment methods:
o Microscope (standard slide: X400; 10 fields of view)
o SQA-V (SQA-V + SQA-V visualization)
o SQA-V visualization system (see table #2).
Immotile cells were analyzed by use of the SQA-V
visualization system.
218 semen specimens contained motile cells and were used
as the basis for the Post Vas visualization method comparison
(Table #5).
Table #3: Precision: Trial #1 and #2 (n=154)
Parameter
Range
Entire Range
5-40
Sperm
41-80
Concentration
>80
Entire Range
10-50
Motility
51-55
>55
Table #4: Mean Values and Precision: Trial #4 (n=246)
Mean
Semen
Parameter
Op1
Op2
Sperm
41.0
40.2
Concentration
Total Motility
54.7
56.9
PR Motility
37.9
39.0
NP Motility
16.8
17.9
Morphology
7.6
Note: Op1 - operator 1; Op2 - operator 2
Table #5: Percentage Motile Cells Detected: Trial #3
Postvasectomy mode
Method Comparison of
218 Samples with Motile
Cells
Revision 21_Sep_2015
th
(Trial #4) guidelines.
Method
SQA-V
Microscope
CV%
CV%
3.1
6.1
5.2
5.9
2.1
5.5
2.5
3.2
5.1
7.2
7.6
10.3
1.5
3.4
6.0
4.1
CV, %
SQA-V
Manual
SQA-V
41.4
11.5
54.9
10.7
36.6
13.3
18.4
27.3
7.6
11.5
27.4
# Samples
% Samples
Motile Sperm
Motile Sperm
Detected
Detected
rd
3.4
5.0
Limitations of method:
7.5
Samples were assessed by different operators using a microscope
and the SQA-V. Inter-operator subjectivity may have affected the
6.8
results of the study.