SECTION IX: Error Messages
ERROR
Messages
1. Stabilization Failed:
2. Self-test Failed:
3. Electronic Noise:
Service Manual 24_Apr_2014
Ensure there is no testing capillary in the measurement compartment.
Remove the SQA-V from sources of electronic noise (cell phones, etc.).
Clean measurement compartment (refer to Appendix).
Reboot the SQA-V without a testing capillary in the chamber:
Turn system OFF then back ON at the main switch on the rear panel.
Press the front panel ON/OFF key to begin Auto-Calibration/Stabilization.
Call technical support if failure recurs.
Ensure there is no testing capillary in the measurement compartment.
Remove the SQA-V from sources of electronic noise (cell phones, etc.).
Clean measurement compartment (refer to Appendix).
Reboot the SQA-V without a testing capillary in the chamber:
Turn the system OFF then back ON at the main switch on the rear panel.
Press the front panel ON/OFF key to begin Auto-Calibration and Stabilization.
Call technical support if this message is displayed again. Prepare for technical
support by printing a copy of the SQA-V internal parameters
Ensure there is no testing capillary in the measurement compartment.
Remove SQA-V from sources of electronic noise (cell phones, etc.).
Clean measurement compartment (refer to Appendix).