1.
Normal Pipetting
PIP
Mode
Standard program. A previously set volume
is aspirated into the pipette tip and then
discharged.
2.
Mixing of Samples
PIPmix
Mode
Program for mixing liquids. The sample is
repeatedly aspirated and discharged.
3.
Reverse Pipetting
revPIP
Mode
Program especially for pipetting liquids with a
high viscosity or vapor pressure, or that tend
to foam.
4.
Pipetting for Electrophoresis
GEL
Mode
Program for loading electrophoresis gels. A
predefined sample volume is aspirated at high,
adjustable speed and then slowly discharged.
5.
Dispensing
DISP
Mode
Program for dispensing liquids. An aspirated
volume is dispensed repeatedly in defined
steps.