BRAND Transferpette -8 Operating Manual page 67

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 39
6.
Mount piston
7.
Reassemble piston
and nose cone
assembly
8.
Mount maifold
9.
Performe
reference run
Insert the cleaned or new piston. Screw on the
piston nut and tighten with the mounting tool
Loosen the nose cones by a half rotation. After-
ward push the nose cone assembly on the piston
unit and fasten it. Then tighten nose cones.
Reassemble the manifold and mount it to the
handle.
Note:
The manifold must be screwed counter-clockwise
into the handle so that it audibly snaps into place.
You must not pull the manifold downward.
Check the instrument for tightness, mobility and
accuracy (see page 63 "Servicing").
Performe the reference run (rEF).
Servicing and Cleaning
67

Advertisement

Table of Contents
loading

This manual is also suitable for:

Transferpette -12