4.
Remove seal
5.
Mount seal
6.
Reassemble
manifold
7.
Performe
reference run
Push the piston unit down to the bottom. Remove
the seal, inspect it and clean or replace if neces-
sary
Note:
The seal will either remain inside the nose cone
or will stay on the piston after the nose cone is
removed.
If required, grease the piston lightly with the
supplied silicone grease. Push the seal on the
piston with its flat side facing up-ward. Use the
mounting tool to mount and tighten the cleaned
or new nose cone.
Reassemble the manifold and mount it to the
handle.
Note:
The manifold must be screwed into the handle
counter-clockwise, until it audibly snaps into
place.
You must not pull the manifold downward.
Check the instrument for tightness, mobility and
accuracy (see page 63 "Servicing").
Performe the reference run (rEF).