Temperature Calibration Verification Test

Use this procedure to verify that your Applied Biosystems
GeneAmp PCR System 9600 still meets the temperature
accuracy specification.

Equipment Required

You will need the following equipment to perform this procedure:
Temperature Verification System, (Part No. N801-0435)
which includes:
o
o
o
o
o

Verifying the Temperature Calibration

Perform the procedure as described in the following steps. We
recommend that you perform this procedure immediately after
you turn on the instrument.
Note
Verification System for a detailed description on operating the digital
thermometer.
1
8-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Digital thermometer with 9V battery installed
RTD probe
Cotton swabs
Light mineral oil
Probe tray
Refer to the Instructions included with your Temperature
If the heated cover is in the forward position, turn the cover knob
completely counterclockwise, then slide the cover back.