Defining Values for Setpoint # 1

In the following steps, you will change some of the default values
for Setpoint #1.
1
2
3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Setpt #1 Ramp 0:00
94.0C
Hold 0:30
Look for the cursor on the display. It appears under the ramp time.
The default time of 0:00 is most commonly used. This causes the
sample to be heated or cooled to the target temperature as fast as
possible.
For this example, we will not change the default ramp time. Press
ENTER to accept 0:00 minutes.
The cursor moves to the temperature entry.
Setpt #1 Ramp 0:00
94.0C
Hold 0:30
Change the temperature to 96.5 degrees C. Press 9, then 6, then 5.
If you make a mistake, press CE (Clear Entry) to erase the value,
then start again. When the value is correct, press ENTER.
Note that the temperature is set to the nearest tenth of a degree. To
enter a whole number such as 94 degrees C, you must specify 94.0
(press 9, then 4, then 0).