Experiment: SBW is set 2.0 nm automatically. Setup other experiment parameters.
b.
c.
Lamp: Single Front is automatically selected. If not, select Single Front and setup other
experiment parameters.
d.
Setup the parameters. See the
Wavelength Program Mode
5. Select Apply and OK after setting up the parameters.
6. Dip the probe in the blank solution and select the Blank icon. Absorbance of selected wavelength
is changed about 0 Abs.
NOTE:
Before measurement, make sure that that bubble does not exist at the mirror side of the probe.
NOTE:
When sample is measured, any light from outside of the bottled should be blocked as it affects the
measurement result.
7. Clean the probe and dip the probe in the sample solution. Select the Sample icon.
8. Input sample name and select OK.
9. The spectrum and result will be displayed. Save or print results as desired.
Lambda 365 UV Express Software Users Guide
.
18
09931262E