Figure 5-1 Input Wiring - Siemens SITRANS F Series Operating Instructions Manual

Ultrasonic flowmeter
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Connecting
5.1 Transmitter Wiring
3. Pull either the AC or DC wires through transmitter case cable gland and into the
4. Locate power plug P12 and loosen P12 connector screws.
5. As per local electric codes, wire input power connector P12 for AC or DC power
Note
Dress cables and make sure cable length is not excessive as to impede proper
replacement of access cover.
20
transmitter case before wiring power connector.
Fuse F1 Type GMA, 2A
Input Power Connector P12
I/O Terminals TB1
J9 RS-232 Port DB9 Connector
Figure 5-1
Input Wiring
depending on power supply provided.
Connector P12
1
2
3
AC
L1
L2/N
GND
S1 Keypad Enable Switch
I/O Input Cables
Input Power Cables
Sensor Cables
DC
POS +
NEG -
GND
Operating Instructions, 02/2012, A5E03086487-05
Wire Color
Black
White
Green
FST020 IP65 NEMA 4X