Clear Logger
If you use the Logger in its default mode, eventually you will use all the memory available for
Logger storage. When this occurs, you will not be able to log more data until you free up the
memory. The [Clear Logger] command erases ALL stored Logger data. Therefore, you
should evaluate the currently stored data and print any valuable information before using this
command.
Note
Saved Sites also consume Logger RAM.
Clearing Logger Memory
1. From the [Est LogTime Left] menu cell, press the <Down Arrow> key.
2. To access the [Clear Logger] option list press <Right Arrow> key.
3. Scroll to [Yes] by pressing the <Up/Down Arrow> keys.
4. To clear the memory press <ENTER> key.
7.5

Operation Adjust Menu Settings

Introduction
The [Operation Adjust] menu becomes available after the meter type and measurement
channel are active. It is recommended that you use it after the sensors are installed and
operating to "fine-tune" the meter's output characteristics.
Each application presents different data display and output requirements due to unique pipe
and liquid conditions. Use the [Operation Adjust] menu to match flow meter operation to the
site. You can set damping controls for the primary flow rate output. You can define a
Deadband, (usually a very low flow rate), below which the flow output will be forced to zero.
You can also select the flow meter response to a continuous Fault condition.