Functions

7.9 Memory Control

Table 7- 8
Pulse Output Values for Gallons (US)
Totalizer Volume Units
Gallons
Gallons
Gallons
K Gallons
K Gallons
K Gallons
M Gallons
M Gallons
M Gallons
Note
IMPORTANT
Totalizer pulse output conditions:
1. FST020 flow meters with firmware version 2.04.04 have an ON pulse width of 200
milliseconds and a maximum output pulse rate of 2.5 Hz. If the instantaneous rate exceeds
2.5 Hz, output pulses will be buffered in memory and will be outputted when the rate drops
below 2.5 Hz.
2. FST020 flow meters with firmware version 2.04.05 have an ON pulse of 10 milliseconds
and a maximum average output pulse rate of 50 Hz. If the instantaneous rate exceeds 50
Hz, output pulses will be buffered in memory and will be outputted when the rate drops
below 50 Hz.
7.9
Memory Control
Introduction
Memory Control is a reference menu that shows the amount of bytes of data memory left.
The data memory capacity depends on the number and complexity of the site setups stored
in memory and the size of the current Datalogger file. The [Memory Control] menu is located
within the [Meter Facilities] menu.
Table 7- 9
Log Memory Left→
Defragment→
90
Totalizer Resolution
000x0000
0000x000
00000x00
000x0000
0000x000
00000x00
000x0000
0000x000
00000x00
Memory Control Menu
Totalizer Represents GAL/Pulse
0.0001 Gallons / pulse
0.001 Gallons / pulse
0.01 Gallons / pulse
0.1 Gallons / pulse
1 Gallon / pulse
10 Gallons / pulse
100 Gallons / pulse
1000 Gallons / pulse
10000 Gallons / pulse
XXXXXXX
No
Yes
Operating Instructions, 02/2012, A5E03086487-05