Installation
Only qualified specialists or persons
trained by Air Techniques may install,
connect, and commission the unit
EN
6

Requirements

US
6.1 Installation/setup room
The room chosen for set up should fulfil the fol-
lowing requirements:
– Closed, dry room
– It should not be a room made for another
purpose (egboiler room or wet cell)
– There should be no major fields of interference
(egstrong magnetic fields) present that can
interfere with the proper operation of the unit
– The required environmental conditions are
met (refer to "Technical Data" in the operating
instructions)
Portable and mobile HF communication appli-
ances can interfere with the effectiveness of
electrical medical devices
Do not stack the unit next to or together with
i
other appliances
If, however, this unit is operated next to other
i
units or stacked with other units, monitor the
unit carefully in the configuration selected in
order to ensure normal operation
WARNING
Risk of explosion due to inflamma-
tion of combustible materials
Do not use the unit in rooms in which
i
combustible mixtures may be present,
eg in operating rooms
18
Radiation protection
Comply with all applicable X-ray protection
i
regulations and X-ray protection measures
Use the prescribed X-ray protection equip-
i
ment
In order to reduce the level of X-ray exposure,
i
we recommend the use of bismuth, lead
shielding or protective aprons, especially for
children and teenagers
The persons operating the equipment must
i
keep away from the X-ray unit while the expo-
sure is being taken The minimum distance re-
quired by the law must be maintained
Children and pregnant women must consult a
i
doctor before recording an X-ray image
No person other than the patient is permitted
i
to be present in the radiation room without X-
ray protection measures In exceptional cir-
cumstances another person may be present
to provide assistance, but this must not be a
member of the surgery staff When the expo-
sure is being taken, make sure that you main-
tain visual contact with the patient and the
unit and keep talking to the patient
The radiation room must be lockable to pre-
i
vent entry by unauthorized persons
If a fault occurs, abort the exposure immedi-
i
ately by releasing the trigger button
Information about electrical connections
Ensure that the electrical connections to the
i
mains power supply are established in accor-
dance with current valid national and local
regulations and standards governing the in-
stallation of low voltage units in medical facili-
ties
Install an all-pole disconnect switch with a
i
contact opening width of at least 3mm in the
electrical connection to the mains power sup-
ply It must be possible to secure the discon-
nect switch so that it cannot be inadvertently
switched back on again