Installation Information; Mojave Lt3 Vacuum Pump Dimensions - Air Techniques MOJAVE LT3 User And Installation Manual

Dry vacuum system
Table of Contents

Advertisement

General. For new installations it is recommended to follow the following guidelines:
 Make sure to install the system in accordance with all local electrical and plumbing codes .
 The suction line should not have any sharp right angle bends and must be sloped a minimum
¼
of
inch for every 10 feet toward the MOJAVE LT3 .
 The drain on the base of the air/water separator must be connected to a vented or an open
floor drain capable of handling 5 gallons in 1 minute . Drain tube size 1 inch .
 The drain line should be a short run with a maximum rise of 3 feet . Avoid any sharp right
angle bends .
 Make sure to install the supplied exhaust vent assembly to the bottom end of the facility vent line .
 The vent should be sloped
of handling vapors and liquids .
 The outside vent must be protected from rain and animals .
 A flexible air exhaust hose is provided to connect to the 1½ inch diameter vent pipe and heat
exchanger . 1½ inch no-hub adapter is provided to secure hose to heat exchanger and pipe .
19.5 in. (50 cm)
16.5 in. (42 cm)
Left Side View
Air Techniques, Inc.
¼
inch per 10 feet towards the pump . Vent lines must be capable
20 in.
(51 cm)
Figure 3. MOJAVE LT3 Vacuum Pump Dimensions

INSTALLATION INFORMATION

24 in. (61 cm)
Front View