X-ray emitter
Nominal current, high voltage generator
Cooling, high-voltage generator
Additional filtration at 50 kV
Integrated filtration at 50 kV
Total filtration at 50 kV
X-ray tube model
Focal spot size as per IEC 60336 X-ray
tube
Anode angle
Anode heat capacity
Pulse/break ratio
Duration of radiation exposure
Maximum current-time product per hour
4.1 X-ray tube performance data
– Maximum deviation of the voltage peak from the displayed value ± 10%
– Maximum deviation of the tube current from the displayed value ± 20%
– Maximum deviation of the exposure time from the displayed value ± 10%
– The device complies with the standards IEC 60601-1, IEC 60601-1-3 and IEC 60601-2-63
– The lowest possible stress factor is obtained with a combination of the settings 60kV and 4mA
2210200497L41 1812V002
mA
mmAl
inAl
mmAl
inAl
mmAl
inAl
mm
mAs
Maximum Rating Charts
DC (Center Grounded)
Constant potential high-voltage generator
Nominal Focus Spot Value: 0.5
25
80kV
20
90kV
100kV
15
10
5
1
2
3
EXPOSURE TIME [s]
4-16 (± 10%, max 75kV16mA,
max99kV10mA)
Automatic monitoring
Shut-off at ≥ 60°C
20 + 30 (autom added for CBCT)
008 + 012 (autom added for CBCT)
28 + 30 (autom added for CBCT)
011 + 012 (autom added for CBCT)
Toshiba D-052SB
in
°
kJ
s
960 (at 75kV/16mA)
70kV
60kV
50kV
5
7
10
20
Product description
08
003
05
002
5
35
1:60 or more
05 - 20
EN
US
9