Product description
Immunity to interference levels by near fields of wireless HF communication devices
Radio service
TETRA 400
GMRS 460
FRS 460
EN
LTE band 13, 17
US
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE band 5
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE bands 1, 3, 4, 25
UMTS
Bluetooth
WLAN 80211 b/g/n
RFID 2450
LTE band 7
WLAN 80211 a/n
Electromagnetic compatibility (EMC)
Interference immunity measurements on supply input
Immunity to interference by rapid transient bursts – AC
voltage grid
IEC61000-4-4:2012
± 2 kV
100 kHz repetition frequency
Immunity to interference, line-line
IEC61000-4-5:2005
± 05 kV, ± 1 kV
Immunity to interference by surges, line-earth
IEC61000-4-5:2005
± 05 kV, ± 1 kV ± 2 kV
Immunity to interference, line-conducted disturbances
induced by high-frequency fields – AC voltage grid
IEC61000-4-6:2013
3 V
015 - 80 MHz
6 V
ISM frequency bands
015 - 80 MHz
80 % AM at 1 kHz
Immunity to interference due to voltage dips, short inter-
ruptions and voltage fluctuations
IEC61000-4-11:2004
12
Frequency band
MHz
380 - 390
430 - 470
704 - 787
800 - 960
1700 - 1990
2400 - 2570
5100 - 5800
Conforms
Conforms
Conforms
Conforms
Conforms
2210200497L41 1812V002
Test level
V/m
27
28
9
28
28
28
9
Need help?