Indicates date of manufacture Indicates protective Earth Ground for the Equipment power switch . MEDICAL ELECTRICAL EQUIPMENT Air Techniques, Inc. WITH RESPECT TO ELECTRICAL SHOCK, FIRE, MECHANICAL 1295 Walt Whitman Road AND OTHER SPECIFIED HAZARDS ONLY Melville, New York, USA 11747- 3062 IN ACCORDANCE WITH UL 60601-1, CAN/CSA C22.2 No.
No applied parts . Protection against ingress of liquids-Ordinary Equipment not suitable for use in the presence of flammable anaesthetic mixture with air or with oxygen or nitrous oxide . All MOJAVE LT vacuum pumps comply with NFPA 99C level 3 requirements. Page 4 Air Techniques, Inc.
Feet Assembly Drain Port Front View Parts Location Sight Glass Vacuum Tube Pump Inlet Manifold Electrical Heat Exchanger Air/Water Vacuum Separator Pump Assembly Assembly Rear View Parts Location Figure 1. MOJAVE LT Key Parts Identification Air Techniques, Inc. Page 5...
Page 6
. Heat Exhaust to Outside Vent Liquids/Solids from the Air/ Water Separator to Sewer Drain Gas/Liquids/Solids from the Treatment Room into Air/Water Separator Figure 2. MOJAVE LT, LT3M and LT5M, Functional Connection Points Page 6 Air Techniques, Inc.
Page 7
Vision Monitor . 19.5 in. (50 cm) 24 in. (61 cm) 16.5 in. (42 cm) 20 in. (51 cm) Front View Left Side View Figure 3. MOJAVE LT, LT3M and LT5M, Vacuum Pump Dimensions Air Techniques, Inc. Page 7...
2. Use maximum internal diameter for the main line when preparing any new installation. Recommended Space Requirement LT3M & LT5M Height 20 in . (51 cm) Width 24 in . (61 cm) Depth 16 .5 in . (42 cm) Page 8 Air Techniques, Inc.
Page 9
. Disconnect boxes should be mounted no more than 3 feet of the installation center line . OVERHEAD INSTALLATION VACUUM LINE - See Plumbing Requirements for MOJAVE LT connection . SUB FLOOR INSTALLATION VACUUM LINE - See Plumbing Requirements for MOJAVE LT connection . Air Techniques, Inc. Page 9...
. Secure with no hub adapters provided by the accessory kit . Heat Exhaust to Outside Vent Liquids/Solids from the Air/ Water Separator to Sewer Drain Inlet Flexible Coupling Connector Figure 5. MOJAVE LT Plumbing Connection Page 10 Air Techniques, Inc.
Page 11
Drip Leg and Flex Tubing Exhaust Vent Assembly 10mm Vent Condensation Drain Port 1" Drain Hose Open Drain Pipe Clamp Connection to Facility Sewer Closed Vented Drain Drain Figure 6. MOJAVE LT Detail Plumbing Connection Air Techniques, Inc. Page 11...
Make the 6 VDC connections shown by Figure 11, View A, for Remote Panel Switch #53202-1 provided by Air Techniques . Make the 24 VDC connections shown by Figure 11, View B, for Remote Panel Switch #53201-1 . When using a switch not provided by Air Techniques, all remote system status indication is lost .
Page 13
BRN 1 Important: Terminals RED & BRN 1 are not used when making the 3-wire connection. View B. 3-Wire Green Indicator Only 24 VDC Remote Switch Installation Figure 9. 3-Wire and 4-Wire Remote Switch Connection Air Techniques, Inc. Page 13...
Self Diagnostic feature to run properly . Unit cannot be turned on/off using a high voltage switch connected to the main power source . Failure to do so will VOID the warranty . Power Circuit Breaker Touch Screen Figure 10. MOJAVE LT Front Panel Controls and Indicators Page 14 Air Techniques, Inc.
Page 15
24V DC Green Indicator Only (See Note a.) Switch Green LED Condition Condition Description Position None Standby, No errors Solid Green Running, No errors Flashing Green – Slow Standby, Error present Green – Fast Running, Error present Optional Remote Control Panel Air Techniques, Inc. Page 15...
Widget - an element of a graphical user interface (gui) that is used to display information during vacuum operation . The Home Screen normally shows the Vacuum Widget while the Motor Frequen- cy, Separator RPMs, Voltage and Temperature widgets are accessed by pressing the Next Navigation Button . Page 16 Air Techniques, Inc.
Page 17
Temperature ( ο F) Widget - Current ambient temperature of room with max temperature limit indicator . Temperature Digital Readout Max Temperature Voltage Indicator Range Indicator Voltage Digital Readout Note: Indicator will turn red when ο room temperature exceeds 105 Voltage Min/Max Temperature Indicator Range Indicator Air Techniques, Inc. Page 17...
Page 18
Scroll Button 2. SETTINGS SCREEN a . INFORMATION Model - Air Techniques model number . ii. SN - Unit serial number . iii. PCB - Indicates control board serial number . iv. Firmware - Indicates latest installed firmware and revision .
Page 19
Run-Time - Shows time that unit has been running (hours) g . ALARM HISTORY Shows the last forty (40) alarms triggered . Push any listed ALARM button to get details of alarm, such as suggested tasks and date alarm was triggered . Air Techniques, Inc. Page 19...
Page 20
Room temperature is too low. iii. Vacuum level is greater than 1.0 inHg over setpoint. Vacuum level is less than 0.5 inHg over setpoint. Line voltage Warning Alert Example Home Screen Showing Warning Example Warning Screen Page 20 Air Techniques, Inc.
Page 21
Separator below minimum speed. Re-enable motor? iii. Motor VFD error detected. Contact Technician. Pressure Sensor Malfunction Pressure Sensor 3 1020011-0 Malfunction Alert Problem Fault Example Pressure Sensor Malfunction Contact Technician Hide Settings Next Home Screen Showing Error Example Error Screen Air Techniques, Inc. Page 21...
.pdf for the complete guide on connecting and installing your Mojave DryStar Monitor to Vision Monitor . The Vision Monitor software that must be installed at the dental practice can be downloaded from this link: https://www .airtechniques .com/en/drivers/ Page 22 Air Techniques, Inc.
Check the inlet filter on the pump assembly for buildup. The filter should be dry and clean. Annually The presence of any liquid or debris means that the installation is incorrect or there is an issue with the air/water separator. Air Techniques, Inc. Page 23...
Page 24
. 5 . With the vacuum pump on and handpiece valves open, aspirate the CleanStream solution from dispenser . 6 . After each cleansing procedure, disconnect the hand pieces and rinse the dispenser . Page 24 Air Techniques, Inc.
ACCESSORIES/OPTIONS Accessories/Equipment Options. The following lists the ordering number and description for accessory components available to maintain the MOJAVE product family . Contact an authorized Air Techniques’ dealer for information . Description Part Number Remote Control Panels with 24V switches:...
Motor VFD error a. Various issues with the VFD a. Check for the VFD error code under detected. Contact can cause this problem the 'Settings -> Info' screen and Technician. contact Technical Support Page 26 Air Techniques, Inc.
Any item returned under warranty, will be repaired or replaced at our option at no charge provided that our inspection shall indicate it to have been defective . Air Techniques, Inc . is not liable for indirect or consequential damages or loss of any nature in connection with this equipment . Dealer labor, shipping and handling charges are not covered by this warranty .
Page 28
For over 50 years, Air Techniques has been a leading innovator and manufacturer of dental products . Our priority is ensuring complete satisfaction by manufacturing reliable products and providing excellent customer and technical support . Whether the need is digital imaging, utility room equipment or merchandise, Air Techniques can provide the solution via our network of authorized professional dealers .