Performance Verification And Maintenance; Extended Self-Test - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Performance
Verification
and
Maintenance
Extended
Self-test
The cardiograph's Extended Self-test, loaded as part of the system
software, tests each major subassembly. The display tests and the
printer test require visual verification of displayed or printed output.
In most cases, a test failure halts the test and a message indicating
the failure appears on the keyboard display. In the case of the tests
that require visual verification, visible failure symptoms appear in the
printed or displayed test pattern. See Chapter 6, "Troubleshooting,"
for a list of self-test failure messages and failure symptoms.
To begin the Extended Self-test, press (On-Standby) to turn on the
cardiograph, then quickly press and release the @FJ and (?J keys
simultaneously. The following display will appear:
Test:
All
Select
Test
Next
The upper line displays the test choice, while the lower line displays
the softkey labels. With the softkeys you step through the test menu,
select tests, and initiate testing.
Select
Selects the test displayed in the message line.
Asterisks (*) appear around the test name to
indicate its selection. If the test is already selected,
pressing this key de-selects the test (the asterisks
disappear). You can select one or more tests in any
combination, but the sequence is fixed.
Test
l3egins running the selected test(s).
Next
Steps forward through the test menu.
Previous
Steps backward through the test menu.
L
ErrorLog
Prints out the stored error log.
To stop or exit Extended Self-test, you must press (On-Standby] to place
the cardiograph in Standby.
M1756A Direct Digital ECG Fax
C-5