Power-On Errors - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Error #
6500
Table 6-1. Power-on Errors
Definition
Corrective Action
Defective ROMs or other related hardware
Replace CPU assembly.
problem on CPU assembly.
6501
6502
6503
Failed to initialize disk controller.
Disk drive
connecting cable bad, disk drive bad, or CPU
assembly bad.
Failed to complete loading of software.
No RAM code found in machine. A.O1.OO
software defect.
1. Replace cable.
2. Replace disk drive.
3. Replace CPU assembly.
Reload software.
1. Reload software.
2. Load software version A.O1.O1
or later.
3. Replace CPU assembly.
6504
RAM cyclic redundancy check (CRC) failed-
Reload software.
RAM code corrupt after loading code.
6505, 6506
6600, 6601
6602
Insufficient memory. M 1700A software loaded
into an M1701A.
Catastrophic
software failure.
Failures in power-on routine.
1. Load proper software.
2. Replace CPU assembly.
Reload software.
1. Cycle power.
2. Reload software.
3. Replace CPU assembly.
6605, 6609, 6610
Missing message table. Disk could be bad.
6606, 6607
Failures in power-on routine.
Reload software.
1. Cycle power.
2. Reload software.
3. Replace CPU assembly.
6612
6613, 6614,
6615, 6616
Wrong message table version.
Reload software.
Software failure.
Reload software.
6990, 6991
Software corrupt.
Reload software.
Note
,I
u
To reload the software on the M1702A
XLe, remove the battery,
then
.
reinstall
the battery
and reconfigure
the unit.
6-8
Troubleshooting

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents