Table 2-3. Global Configuration
Parameters
(continued)
Parameter
Setting Choices
Comments
Auto Frequency Response?
.15-150 Hz
W = Baseline Wander filter. See
.05-150 Hz
Page Writer
XLi
Physician's
Guide for
.5-150 Hz W
more information.
.15-40 Hz
.5-40 Hz W
.05-100 Hz
.15-100 Hz
.5-100 Hz W
Manual Frequency Response?
.05-40 Hz
W = Baseline Wander filter. See
.5-100 Hz W
Page Writer
XLi
Physician's
Guide
for
.05-100 Hz
more information.
.5-150 Hz w
.05-150 Hz
.5-40 Hz W
Line Power Filter?
Line Power Frequency?
Filter Key?
Off
On
60 Hz
50 Hz
BaselineWander
& Artifact
BaselineWander
Artifact
A software filter that removes power
line (AC) interference.
No default-selected
during software
installation.
Determines which filter combinations
are active when (Filter) key is on.
.
Auto print speed?
Default Auto?
Default Manual?
25 nun/set
Default Auto ECG print speed; can be
50 mm/xc
overridden from the front panel.
3x4
Default format for Auto ECGs.
3x4,lR
I
Determines row k column notation for
3x4JR
I, II, III
ECG. Rhythm strip leads also selected
6x2
here ("R." means rhythm strip).
3
Default format for Manual ECGs.
6
Selects number of leads on Manual
12
ECG.