Installation
and
Configuration
Connecting
the Patient
1. Connect the patient data cable to the ECG input jack on the back
Module
of the cardiograph.
2. Connect the other end of the patient data cable to the patient
module as shown in Figure B-l.
3. Connect the leadwires to the patient module. Match the color
coding on the leadwires to the color coding on the patient module.
4. Install shorting plugs in the patient module in those positions
without leadwires.
Figure B-l. Connecting
the Patient Module.
A. Leadwires
B. Patient Data Cable
Note
.I
v
The M1754A SAECG patient module can acquire both SAECG and
standard ECGs. To get the best results for SAECG, remove the LL
and LA leadwires and install shorting plugs in the LL and LA inputs.
B-2
M1754A Signal Averaged
ECG (SAECG) System