Maintenance; Codemaster 100; Operation Checks; Every Shift - HP Codemaster 100 User Manual

Portable defibrillator/monitor
Table of Contents

Advertisement

Maintenance

13

CodeMaster 100

13

Operation Checks

These checks are intended to briefly verify the proper operation of the HP
CodeMaster 100. Regularly perform a test routine incorporating the following
checks along with visual inspection of all cables, paddles, and controls.

Every Shift

Perform the following checks every shift:
o
Make sure that a battery is charging in the battery support system.
o
Turn on the HP CodeMaster 100. If the system displays the LOW
BATTERY message, replace the battery with a charged battery.
o
Check for adequate thermal paper in the recorder.
o
Check for ECG leads, electrodes, and adequate electrolyte medium.
If you have an external power module, perform these additional checks every
shift:
o
Unplug the AC or DC external power module. Turn on the HP
CodeMaster 100. If it displays the LOW BATTERY message, replace the
battery with a charged battery. Turn off the defibrillator.
o
Plug in the external power module. Make sure that the power indicator
on the external power module appears and that the external power
indicator on the HP CodeMaster 100 appears.
o
Remove the battery from the HP CodeMaster 100. Make sure that the
charging indicator on the external power module disappears. Replace the
battery.