Power Supply And Battery Specifications - HP PageWriter XL M1705B Service Manual

Cardiographs, mobile cart, telecommunications package
Table of Contents

Advertisement

Table 1-5. Power Supply and Battery Specifications
Parameter
Specification
AC line frequency
50 and 60 Hz
100 V setting
90-110
V AC
120 V setting
108-132V
AC
220 V setting
198-253
V AC
240 V setting
216-269
V AC
Battery voltage
18 V DC, 6.5 Ahr
Battery recharge time (from 70% down)
to 90% capacity
8 hours typical
to full capacity
16 hours typical
Battery capacity, M1700A with preview
without storage
30 ECGs (40 typical)
with storage
30 ECGs (40 typical)
continuous rhythm (Manual mode)
20 minutes (30 typical)
continuous idle
2 hours (3 typical)
Battery capacity, M1700A without preview
without storage
50 ECGs (70 typical)
with storage
45 ECGs (60 typical)
continuous rhythm (Manual mode)
30 minutes (40 typical)
continuous idle
4 hours (6 typical)
Battery capacity, M1701A
without storage
55 ECGs (75 typical)
with storage
50 ECGs (70 typical)
continuous rhythm (Manual mode)
30 minutes (40 typical)
continuous idle
4 hours (6 typical)
Battery capacity, M 1702A
55 ECGs (75 typical)
continuous rhythm (Manual mode)
30 minut,es (40 typical)
continuous idle
4 hours (6 typical)
Battery life
AC and battery operation
Typically
24 months
battery only operation
Typically
12-18 months
Memory data retention time
with fully charged battery
Typically
1 month
after "Low battery shutdown"
Typically
5 days
h/Iodem power
9 V AC for Hayes@ modem
9 V DC for Worldport
249BTM Fax/Data
modem
Ground leakage current,
at 120 V
Less than 100 /iA
at 220 V
Less than 500 /LA
Dielectric withstand voltage
1.5 kV RMS across line/neutral
to Earth for 1 min
Introduction
l-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents