OPERATION

2.2.2 Lid Lock

The lid must only be closed slightly. An electromagnetic lid lock closes the lid, the word "open" (M2)
will no longer be displayed.
As a sign that the centrifuge is ready for starting, appearing in the display, "rpm/rcf"(A-1), the word
"close" (M1). Simultaneously, the word "rotor" (M3) appears, as well as the code number of the
rotor, which is in the centrifuge, i. e. " 221.40" (M4), along with all rotor specific data, for example:
max. speed, acceleration etc., are available.
For all number marked text, please refer to Figure16.
ATTENTION: Don't grip your fingers between the lid and the device or the locking

2.3 Pre-Selection

2.3.1 Pre-Selection of Speed / RCF-Value

Selecting the key, "rpm/rcf" (4), pre-selection is activated. By pressing the key once, the word "rpm"
(M5) flashes. By pressing the key again, the pre-selection of the centrifugal forces can be chosen.
The flashing word, "rcf" (M6), will appear. The desired values can be selected, with the adjustable
knob (1). In the display (A-1), the regulated value is shown permanently: before, during and after the
run.
For all number marked text, please refer to Figure17.
Figure17
The speed is adjustable between 200 rpm and the max. speed of the centrifuge respectively of the
rotor. It is the same with the pre-selection of the RCF-Value. The setting range is between 20 xg and
the maximum permissible centrifugal force of the centrifuge respectively rotor.
The max. permissible speed of the OLEUM with rotor 221.40 V20 is 3,000 rpm or 2,213 xg.
ATTENTION:
Please check the maximum permissible revolutions of your test tubes! (Producer
Indication).
16
mechanism, when closing the lid!
OLEUM_V1.19_eng
© Hermle Labortechnik GmbH