Operation; Mounting And Loading The Angle Rotor; Installation Of Rotors; Loading Swing Out Rotor - Hermle OLEUM Operating Manual

Oil test centrifuge
Table of Contents

Advertisement

2. OPERATION

2.1 Mounting and Loading the Angle Rotor

2.1.1 Installation of Rotors

Clean the drive shaft and the collet with a clean, grease-free piece of cloth. Place the rotor onto the
drive shaft, (see Figure12). Please be sure that the rotor is fully installed onto the motor shaft.
Figure12
Hold the rotor with one hand and secure the rotor to the shaft, by turning the fixing nut clockwise
(see Figure 12). Tighten the fixing nut with the provided Allen key, (see Figure13).
Figure13
ATTENTION: For safety, always ensure that the rotor fixing screw is tightened before
each run, (see

2.1.2 Loading Swing Out Rotor

Loading of the buckets must be done in accordance to Figure15.
It is allowed to operate e.g. a 4-place-rotor with 2 loaded buckets only. The loaded buckets must be
opposite of each other. Make sure that the unloaded buckets are placed inside the rotor.
In principle, the swing out rotor is not allowed to be operated, until all buckets are placed inside the
rotor. The bolts of the rotor must be greased with the HERMLE Lubricating grease
(Order No. 38-5656). The sample tubes have to be filled evenly, by eye, and set into the buckets.
The weight difference of the loaded buckets should not exceed 1.0 g.