Section 7: Preventive Maintenance

Weekly Preventive Maintenance

Important
The operator of the PROCESSOR should perform the weekly preventive maintenance procedures in this section.
2 EVAPORATION
COVERS
WATER
RESERVOIR
WATER DISCONNECT
[1] De-energize the PROCESSOR.
[2] Remove the WET SECTION COVER.
[3] Disconnect the WATER RESERVOIR or YOKE from the CROSSOVERS.
[4] Remove the following parts:
a. EVAPORATION COVERS
b. all CROSSOVERS
c. WASH RACK
[5] Rinse the 3 CROSSOVERS and the WASH RACK with warm water, 44°C (110°F) or less. Wipe the
ROLLERS and the GUIDE SHOES with a damp, soft, synthetic sponge.
[6] Allow the DETECTOR CROSSOVER to air dry before you install it in the PROCESSOR.
[7] Check that all the ROLLERS on all the CROSSOVERS and on the WASH RACK rotate freely.
[8] Check the squareness of the CROSSOVERS and the WASH RACK. See the procedures on Page 8.
[9] Check that the CROSSOVER TROUGHS are clean, are without cracks, and are installed correctly.
[10] Wipe any deposits from the processing section of the PROCESSOR.
[11] Install the following parts. Check that each assembly seats firmly.
a. WASH RACK
b. all CROSSOVERS
28
WASH
DRYER RACK
RACK
WET SECTION
COVER
FIXER/
WASH
CROSSOVER
DETECTOR
CROSSOVER
DEVELOPER
RACK
DEVELOPER/FIXER
CROSSOVER
FIXER
RACK
H104_0586HCA
H104_0586HA
May 1996 – SM3058-3

Advertisement

Table of Contents
loading

This manual is also suitable for:

X-omat multiloader 300

Table of Contents