[8] On all RACKS, check that:
(a) all ROLLERS rotate freely
(b) the ROLLERS are clean and smooth
(c) the GEARS do not have excessive wear
(d) the BEARINGS do not have excessive wear
(e) the E-RINGS and SPRINGS are not broken, missing, or corroded
[9] If necessary, install new parts.
[10] Check the RACKS for squareness. If necessary, see the procedure on Page 8.
[11] Check that the SLOTS in the AIR TUBES are clean. If necessary, clean the AIR TUBES with a BOTTLE
BRUSH and rinse with water.
[12] Wipe any deposits from the processing section of the PROCESSOR.
[13] Change the DEVELOPER FILTER.
[14] Allow the DETECTOR CROSSOVER to air dry before you install it in the PROCESSOR.
[15] For PROCESSORS with a WATER RESERVOIR, clean the WATER RESERVOIR. Check that the holes in
the WATER RESERVOIR are free of biological growth and other obstructions.
[16] Install the RACKS.
[17] Install the following parts. Check that each part seats firmly.
a. CROSSOVERS
b. WATER RESERVOIR or YOKE and WATER DISCONNECT
c. EVAPORATION COVERS
d. WET SECTION COVER
[18] Clean the REPLENISHMENT STRAINERS that are between the REPLENISHMENT TANKS and the
REPLENISHMENT PUMPS.
[19] Check the water supply FILTER. If necessary, change the FILTER.
Caution
• Any obstruction such as biological growth, small bends, or an upward sloping WASH DRAIN HOSE can cause
wash water to return to the WASH TANK causing an overflow of water.
• Concentrations of chlorine bleach higher than 10% or allowing the chlorine bleach to remain in the
PROCESSOR for longer than 30 minutes can cause damage to the PROCESSOR.
[20] Check the WASH DRAIN for biological growth, small bends, or other obstructions. To remove biological
growth do the steps below:
(a) Fill the WASH TANK and DRAIN with a 10% solution of chlorine bleach.
(b) Wait a minimum of 20 minutes and a maximum of 30 minutes.
(c) Drain the chlorine bleach solution and rinse the TANK and DRAIN thoroughly with water before you
return the PROCESSOR to normal use.
[21] Energize the PROCESSOR.
[22] Process a test film and check for correct operation of the PROCESSOR.
30
May 1996 – SM3058-3