Before You Call Service
The 220 Indicator has been designed to provide you with years of trouble-free operation. In spite
of this, troubles sometimes happen. Before calling for service assistance you should make some
initial checks to verify that a problem does exist. The following describes several types of
symptoms along with suggested remedies.
PROBLEM
Display does not turn on
Incorrect weight displayed
Indicator will not display weight
CALIBRATION SEAL INSTALLATION
If your 220 Weight Indicating Instrument is used in a commercial application it must be tested and
sealed by your local weights and measurements official. The 220 is designed to accept a lead and
wire security seal to prevent unauthorized access to the calibration adjustments. Refer to the
Figure No. 9 for details on the installation of these seals.
TROUBLESHOOTING, Cont.
Is the AC power cord fully inserted into the wall
receptacle? Check wall receptacle for proper AC power.
Try another electrical appliance in the same receptacle,
does it work? Check the circuit breaker. Has there been
power failure?
Has the instrument been calibrated? Insure that the
scale platform isn't touching an adjacent object. Check
the load cell connector wiring. If using four (4) wire load
cells, insure the sense lead jumpers (J4 & J5) are
installed. Have proper operation procedures been
followed?
Refer to Error Codes section and make certain that the
OVER CAP message is not displayed. If so, and scale is
not loaded, perform the calibration sequence.
Figure No. 9
28
POSSIBLE SOLUTIONS