Procedure

3.8.1 Transferring the adapter position

Objective
The transfer device is used to transfer the adapter position from the check prosthesis to the definitive prosthesis.
The check socket is then filled with plaster and adapted to the plaster model.
Required tools and materials
743A160 transfer device, oscillating saw, tape measure, 716Y3 half-round file, 716Y4 round file, 649G22=180
grit cloth

3.8.2 Vacuum forming

Objective
A layer of ThermoLyn PETG is vacuum formed over the plaster model as the inner wall of the prosthetic socket. The
ThermoLyn PETG reduces the risk of leaks in the prosthetic socket.
Required tools and materials
616T483=3 ThermoLyn PETG
20 | Ottobock
Remove all parts used for the check socket of the liner
FitKit.
INFORMATION: If more than three parts from the
liner FitKit were used for the check socket, a new
check socket should be made.
Transfer the adapter position of the check prosthesis to
the transfer device.
Fill the check socket with plaster to secure the vacuum
pipe.
Remove the plaster model from the check socket.
Prepare the plaster model for the lamination process.
Reduce the plaster model according to the parts of the
liner FitKit used:
Thick residual limb sock: Reduce the circumfer­
ence by 5 mm
Spot: Reduce the plaster model locally by 2 mm
Complete the vacuum forming process with ThermoLyn
PETG.
Cut off material protruding over the model with a knife.
TIP: Further information on vacuum forming with PETG
can be found in the 646T4=1.0 technical information.