Otto Bock Harmony P3 Technical Information page 11

Fabricating a transtibial prosthesis
Hide thumbs Also See for Harmony P3:
Table of Contents

Advertisement

Required tools and materials
645C2=* copying pen, tape measure, measurement sheet and reduction table (see appendices), 716Y3 half-
round file, 716Y4 round file, 649G22=180 grit cloth
Harmony System
Remove the plaster cast from the residual limb, fill with
plaster, insert a reinforcing bar and allow the plaster
model to cure.
Identify and mark reference points for the circumferen­
tial and length measurements on the plaster model.
Determine the circumferential and length dimensions.
Record the dimensions on the measurement sheet.
Divide the plaster model into four sections (anterior,
posterior, medial and lateral) and mark.
Mark the transitions between convex and concave
shapes with horizontal lines.
Reduce the distal end of the model by 5 mm to 8 mm
without changing the shape.
If there is overhang of soft tissue, more reduction may
be necessary.
Reduce the circumference of the plaster model globally
by 4% or dependent on the residual limb shape and the
tissue condition.
Reduce the circumference of the plaster model evenly,
dependent on the residual limb shape and the tissue
condition:
MPT: Reduce by 2% to 2.5%
Area a: Reduce by 3% to 4%
Area b:
Reduce globally by 3% to 5%
Muscular residual limb with firm soft tissue coverage
Cylindrical residual limb: 4%
Conical residual limb: 5%
Soft residual limb with easily movable soft tissues
Cylindrical residual limb: 3% to 4%
Conical residual limb: 4%
Procedure
Ottobock | 11