Cleaning & Maintenance
Before connecting the pump to a patient, and periodically during use, clean
the unit as instructed below.
• Always turn the pump off and remove the battery before
cleaning.
• Always unplug the charger from AC power before cleaning.
• Do not clean the pump or charger with chemicals such
as Xylene, Acetone, or similar solvents. These chemicals can
cause damage to plastic components and paint. Use a lint-
free cloth with hydrogen peroxide 10% or isopropyl alcohol
70%.
• Do not steam autoclave, ethylene oxide sterilize or immerse
any part of the pump or charger in any fluid
Periodic maintenance is recommended every 1 year. Periodic maintenance is
designed to ensure the pump's accuracy and detect and repair any potential
pump inconsistencies prior to their occurrence in the field. Refer to the direc-
tions for use for information about periodic maintenance.
There is no additional time required for the unit to equilibrate once removed
from minimum to maximum storage temperatures. Ensure to operate the
pump within the specified operating range.
The expected service life of the pump and charger is 3 years.
The rating of the pump is IP33. This is the moisture protection rating and
indicates the degree of particle and water ingress protection.