To drain and fill the ultrasonic wash supply bottle:

!
WARNING
WARNING! You will be handling ethanol in this procedure. For safe handling procedures, refer
to the MSDS provided by your ethanol supplier. At a minimum: 1) wear protective clothing
(laboratory gloves, eye protection, and laboratory coat), 2) have adequate laboratory ventilation,
3) keep away from ignition sources, 4) and have proper fire extinguishing devices/solutions readily
accessible.
1.
Prepare at least 100 mL of a 50:50 solution by volume of Type 1, > 18.2
100% ethanol.
2.
Go to the Maintenance screen (tap the Back button
MAINTENANCE button
3.
Tap SONICATOR SOLUTION Drain on the Maintenance screen
This drains the ultrasonic wash. An on-screen message displays the time remaining
for the operation. Wait for draining to complete.
Figure 4.8 Draining the Ultrasonic Wash
4.
Tap SONICATOR SOLUTION Fill on the Maintenance screen
This moves the pin array head so that you can access the ultrasonic wash supply
bottle.
Figure 4.9 Maintenance Screen
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
Chapter 4 Maintenance Procedures
on the Main Menu).
water and
on any screen and tap the
(Figure
4.8).
(Figure
4.9).
USG-CUS-059 Rev05
95