Opening Or Closing The Main Door - Agena BIOSCIENCE MassARRAY Nanodispenser RS1000 User Manual

Table of Contents

Advertisement

Main door and front
access panel
!
CAUTION
Opening the main door disengages the safety interlock and aborts the dispensing run in progress.
See
page 15
for more details on the enclosure safety interlocks.

Opening or closing the main door

To open the main door, lift the door fully open using the handle
door is approximately half-way open, the gas springs will assist lifting the door. The door
remains in the open position until you close it. The safety interlock message is displayed
while the door is open
(Figure
To close the main door, grasp and hold the handle as you close the door. Tap the HOME
button in the Safety Interlock message after the door is closed.
!
WARNING
WARNING! Do not place hands in the main door pathway when closing it. When lowering the
main door, do not let go of the handle until the door is completely closed. The gas springs support
the door as it is lowered, but stop providing support when the door is almost closed. If you do not
hold onto the door handle, the door may slam closed.
Figure 1.8 Nanodispenser RS1000 front side
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
1.9).
Handle
Chapter 1 Introduction
(Figure
1.8). Once the
USG-CUS-059 Rev05
9