Appendix A Changing the Pin Format
128
NOTE
If you are going to use the standard 24-pin format, no vacuum cover should be installed. Skip step
11
and proceed to installing the calibrant reservoir (step
11.
Place the vacuum cover on the dry station:
a. Select the proper vacuum cover for the installed insert.
!
CAUTION
It is important to use the correct vacuum cover for the pin installed insert. Otherwise, pins will not
be washed properly and samples may become cross-contaminated.
Figure A.19 Vacuum Covers
6-pin vacuum cover
(6 holes)
Alignment pin
Alignment pins
b. Orient the vacuum cover so that there are two alignment pins forward and one pin
at the back.
c. Put the vacuum cover on the dry station so that the alignment pins (on the cover
bottom) fit into the wells and L-shaped channel indicated in
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
12
on
1-pin vacuum cover
(1 hole)
Alignment pin
Alignment pins
page 129
).
Figure
A.20.
USG-CUS-059 Rev05