Preventive Maintenance Schedule
6.3
6.3.1
6.3.2
6 - 2
To Purchase, Visit
Avobus.com
Most simulators (properly maintained) will provide reasonably accurate and consistent
results, but may not produce the exact results when connected to similar monitors from
different manufacturers. However, the consistent performance of a simulator is a good
indicator of any change in the performance in the monitor.
*
For example, an accurate mercury column for pressure verification.
Preventive Maintenance Schedule
The following preventive maintenance steps are required for continued satisfactory
performance and safety of the Accutorr Plus and optional accessories. Inspections and
replacement of consumable supplies and accessories that are subject to normal wear, must
be accomplished at least as frequently as the inspection of the host unit. Read the Warranty
statement in the Operating Instructions for description of warranty conditions.
The suggested minimum maintenance schedule is based on normally expected wear and
tear of system and components. The intervals may be shortened at the user's discretion
when conditions of use may warrant. In case of internal instrument failure, component
replacement, and possible or actual instrument damage, it is strongly recommended that
complete performance and safety verification be conducted.
Mechanical and Physical Visual Inspection (One Year Interval)
Check the following items for wear and physical damage: Repair or replace as needed.
• Outer case of Accutorr Plus, Recorder and Temperature module.
• Patient connected accessories such as blood pressure cuffs and hoses and related
connectors. Check the SpO
probes. Verify and replenish the disposable items as needed. (i.e.,: probe covers,
recorder chart supply).
• Power cords, external safety ground connections and communication cables.
Electrical Safety and Performance Checks (One Year Interval)
• Line power and battery power indicator lights functional in the proper modes.
• Line power and safety grounds are intact and within specifications.
• Non-invasive blood pressure section is functional within specifications.
• SpO
functions are operational and within specification.
2
• Temperature functions are operational and within specifications.
• Recorder is functional and documents available stored data.
• Check recorder paper path and clean as required.
• Error codes, failure messages, end user reported anomalies are collected and analyzed
for reduced performance trends. Identify and correct causes as warranted.
• Perform ground resistance and chassis leakage tests.
or call
1-800-674-3655
sensor and cables. Check the temperature cables and
2
0070-10-0691
Preventive Maintenance
Accutorr Plus Service Manual