Specifications
2.3
Physical Characteristics
Size (maximum):
2.4
Environmental Characteristics
Operating
Temperature:
Humidity:
Altitude:
Shock and Vibration:
Shipping:
Storage
Temperature:
Humidity:
Accutorr Plus Service Manual
Main Unit:
Recorder Module:
Predictive Module:
Weight:
0070-10-0691
19 cm(W) x 26.93 cm(H) x 20.83 cm (D)
7.5" (W) x 10.6" (H) x 8.2" (D)
5.33 cm(W) x 23 cm(H) x 11 cm (D)
2.1" (W) x 9" (H) x 4.25" (D)
5.7 cm(W) x 15.9 cm(H) x 11.8 cm (D)
2.25" (W) x 6.25" (H) x 4.63" (D)
<4.95 kg (11 pounds), depending on configuration
10°C to 40°C, 50°F to 104°F (Accutorr Plus
& Recorder)
10°C to 32°C, 50°F to 90°F (Predictive
Thermometer)
15 to 90% max, non-condensing.
1013 hPa to 697 hPa (0 to 10,000 ft.)
Meets IEC 60068-2-27 for shock with peak
acceleration of 15g, 11 mSec duration, half sine.
Meets EN 60068-2-64 for random vibration
with frequencies of 10 to 2000 Hz, resolution of
10Hz, 10 minutes per axis, acceleration spectral
density of:
10 Hz to 100 Hz : 1.0 (m/s2)2/Hz
100 Hz to 200 Hz : -3 dB per octave
200 Hz to 2000 Hz : 0.5 (m/s2)2/Hz
Meets ISTA Test Procedure 1A (less than 100 lbs)
-15°C to +60°C
+5°F to 140°F
10 to 95%, non-condensing